JTR-8012 intelligent rail type temperature transmitter
Classification:
Keywords: JTR-8012 intelligent rail type temperature transmitter
E-mail:
Product Description
Input type: 2/3/4 wire PT100 and tungsten rhenium/B/R/S/N/K/J/T/E type thermocouple
Output mode: 0~10V (can be adjusted). RS485 (optional)
Output resolution: 1mv
Data update cycle: 0.5 seconds
Thermocouple cold junction compensation accuracy (AT 25 ℃):(max) ± 0.1 ℃
Operating environment:-30~70 ℃ 0 ~ 95% relative humidity
Special function: can be matched with the sensor, to achieve ultra-high precision measurement.
When the thermocouple is input, the temperature compensation element can be externally connected to realize instant compensation.
Sensor open indicator alarm.
Electrical isolation: power-input-output/RS485
Isolated from each other, isolation voltage 600V
Power supply: 18 ~ 36VDC (other range can be customized)
Accuracy (AT 25 ℃ & AT 0.5 maximum sampling rate):
Enter model number | Range (℃) | A/D conversion accuracy | D/A conversion accuracy | Influence of 1 ℃ of environment change (temperature drift) |
PT100 | -200~500 | ±0.05℃ | ± 0.05% of range | 0.003 ℃ |
Type B thermocouple | 300~1820 | ± 1.40 ℃ | ± 0.05% of range | 0.030 ℃ |
S-type thermocouple | 0~1768 | ± 0.65 ℃ | ± 0.05% of range | 0.025 ℃ |
N-type thermocouple | -200~1300 | ± 0.35 ℃ | ± 0.05% of range | 0.019 ℃ |
K-type thermocouple | -180~1372 | ± 0.25 ℃ | ± 0.05% of range | 0.018 ℃ |
J-type thermocouple | -180~760 | ± 0.20 ℃ | ± 0.05% of range | 0.006 ℃ |
T-type thermocouple | -200~400 | ± 0.20 ℃ | ± 0.05% of range | 0.009 ℃ |
Type E thermocouple | -50~1000 | ± 0.17 ℃ | ± 0.05% of range | 0.006 ℃ |
digital/analog conversion |
|
|
| 0.002% Span |
Note 1: The digital/analog conversion accuracy and its temperature drift given in the table are for the output range of 0~10V.
Other range calculation methods: digital/analog conversion accuracy = 0.05*10/range; Digital/analog conversion temperature drift = 0.002*10/range.
For example, 1~5V output: digital/analog conversion accuracy = 0.05*10/(5 - 1) = 0.125
Note 2: Total digital accuracy of thermocouple measurement: analog/digital conversion accuracy 0.1 ℃ (cold end compensation accuracy)
Note 3: Total accuracy of transmitter: analog/digital conversion accuracy. For calculation method, please refer to the previous "About accuracy"
Overall accuracy after using sensor matching function (transmitter sensor)
Enter model number | Range (℃) | Accuracy of analog/digital conversion (℃) |
PT100 | -50~200 | ±0.05 |
S-type thermocouple | 0~650 | ±0.70 |
650~1100 | ±1.50 | |
N-type thermocouple | 0~650 | ±0.40 |
650~1100 | ±1.90 | |
T-type thermocouple | -200~-70 | ±0.40 |
-70~400 | ±0.20 |
Note 1: The sensors listed in the table are the most commonly used, stable and representative sensors in each temperature range. Consult the manufacturer for details of accuracy in a wider temperature range.
Note 2: The thermocouple matching function is a patented technology and must be completed before leaving the factory.
Note 3: The accuracy of the thermocouple in the table includes the cold junction compensation precision.
More products
Online Message